This table is a quick reference point to identify basic system specifications, system requirements, and performance figures for the LAUREATE® World Phaco System.
Dimensions
Height: 24.1 cm (9.5 inches)
Width: 53.3 cm (21.0 inches)
Depth: 40.5 cm (16.0 inches)
Weight
Unpacked: 17.3 kg (38 pounds)
Packed: 27.3 kg (60 pounds)
Environmental Limitations
| Operating | Non-operating |
| Altitude: | |
| 3,048 meters | 12,191 meters |
| (10,000 feet) | (40,000 feet) |
| Temperature: | |
| 10°C to 35°C | -40°C to 60°C |
| (50°F to 95°F) | (-40°F to 140°F) |
| Relative Humidity: | |
| 10% to 95% without condensation | 10% to 95% without condensation |
Electrical Requirements
100 - 120 VAC, 50/60 Hz
220 - 240 VAC, 50/60 Hz
Remote Control
Method: Infrared
Channels: 4
Batteries: AAA (3)
Maximum Input Current
6 A
Protection Against Electric Shock
Class 1
Classification of all Applied Parts
Type BF
Data Card
MMC (Multimedia Card), or SD (Secure Digital) 32 Mb min
Phacoemulsification
Submodes: Pulse, Continuous, Burst
Tip Stroke @ 100%: 88.9 ± 27.0 µm (.0035 ± .0005 in)
Resonant Frequency: 38.0 ± 1.9 kHz
Pulse Rate Range: 0-100 pulse/second (depending on % Time On)
Burst Length: 5 to 500 ms
U/S Power: U/S power is defined as being proportional to ultrasound displacement of the phaco tip
Anterior Vitrectomy
Submodes: Cut I/A, I/A Cut
Cut Rate: 10 to 1500 cuts/minute
Coagulation
10 Watts max, 75 ohm load
110 Vpp @ 1.5 MHz ± 75 kHz, 75 ohm load
129 Vpp @ 1.5 MHz ± 75 kHz, no load
Vacuum & Aspiration @ Sea level:
Vacuum: 0 to 650 mmHg (0 to 86 kPa)
Aspiration: Up to 60 cc per minute (mL/minute)
Power IV Pole on Mobile Cart (optional):
Height Range: 30 to 102 cm